A&D SW-15KS Benutzerhandbuch

Seite von 46
32
 
Command examples    (“_” indicates “Space” (20H).) 
The examples below are for the function setting “
aCk 1
 “ (Reply to commands). 
‰
 Request the weight data.   
Command 
Q C
R
L
F
 
 
Reply 
S T ,  + 0 0 1 2
.
3 4 5 _ k g C
R
L
F
Stable positive data 
 
U  S , 
+  0  0  0 7
.
8 9 0 _ k g C
R
L
F
Unstable positive data 
 
O  L
+  9  9  9 9
.
9 9 9 _ k g C
R
L
F
‘E’ display 
 
‰
 Zero the scale. (No reply for the function setting “
aCk 0
 “.) 
Command 
Z C
R
L
F
 
   
 
Reply 
Z C
R
L
F
 
 
The scale is in a condition that zero operation is possible. 
   
‰
 Tare the scale. (No reply for the function setting “
aCk 0
 “.) 
Command 
T C
R
L
F
 
   
 
Reply 
T C
R
L
F
 
 
The scale is in a condition that tare operation is possible.
 
 
‰
 Switch the weighing unit. (No reply for the function setting “
aCk 0
 “.) 
Command 
U C
R
L
F
 
   
 
Reply 
U C
R
L
F
 
 
Switch the weighing unit to the next weighing unit. 
 
‰
 Five-level comparator mode…Send the current HIHI limit value. 
Three-level comparator mode…Send the current HI limit value. 
Command 
? H 2 C
R
  L
F
 
 
Reply 
H 2
,  + 0 0 0 4 0 0 C
R
L
F
 
 
‰
  Five-level comparator mode…Send the current HI limit value. 
Three-level comparator mode…Not used 
Command 
? H 1 C
R
  L
F
 
 
Reply 
H 1
,  + 0 0 0 3 0 0 C
R
L
F
 
‰
  Five-level comparator mode…Send the current LO limit value. 
Three-level comparator mode…Not used 
Command 
? L 1 C
R
  L
F
 
 
Reply 
L 1
,  + 0 0 0 2 0 0 C
R
L
F
 
‰
  Five-level comparator mode…Send the current LOLO limit value. 
Three-level comparator mode…Send the current LO limit value. 
Command 
? L 2 C
R
  L
F
 
 
Reply 
L 2
,  + 0 0 0 1 0 0 C
R
L
F