A&D SW-60KM Benutzerhandbuch

Seite von 46
33
 
 
‰
  Five-level comparator mode…Set the HIHI limit value. 
Three-level comparator mode…Set the HI limit value. 
(No reply for the function setting “
aCk 0
 “.)
 
Set the six-digit value excluding the polarity and decimal point.
 
Command 
H 2
,  +  0  0  0 4 0 0 C
R
L
F
 
Reply 
H 2
,  + 0 0 0 4 0 0 C
R
L
F
 
‰
 Five-level comparator mode…Set the HI limit value. 
Three-level comparator mode…Not used. 
(No reply for the function setting “
aCk 0
 “.)
 
Send the six-digit value excluding the polarity and decimal point.
 
Command 
H 1
,  +  0  0  0 3 0 0 C
R
L
F
 
Reply 
H 1
,  +  0  0  0 3 0 0 C
R
L
F
 
‰
 Five-level comparator mode…Set the LO limit value. 
Three-level comparator mode…Not used. 
(No reply for the function setting “
aCk 0
 “.)
 
Send the six-digit value excluding the polarity and decimal point.
 
Command 
L 1
,  +  0  0  0 2 0 0 C
R
L
F
 
Reply 
L 1
,  +  0  0  0 2 0 0 C
R
L
F
 
‰
 Five-level comparator mode…Set the LOLO limit value. 
Three-level comparator mode…Set the LO limit value. 
(No reply for the function setting “
aCk 0
 “.)
 
Send the six-digit value excluding the polarity and decimal point.
 
Command 
L 2
,  +  0  0  0 1 0 0 C
R
L
F
 
Reply 
L 2
,  +  0  0  0 1 0 0 C
R
L
F
 
Replies to the commands other than examples above when the function setting “
aCk 1 
“ is 
selected
.
 
‰
 The scale is not in a state where a command could be executed. Then, the scale will reply “I”.   
Command 
Z C
R
L
F
 
   
 
Reply 
I C
R
L
F
 
 
The scale is not in a condition that zero operation is possible.
 
‰
 Command does not exist for the scale. Then, the scale will reply “?”.   
Command 
B C
R
L
F
 
 
 
 
Reply 
? C
R
L
F
 
 
The scale received an undefined command.
 
 
‰
 When the function setting “
aCk 0
 “ is selected, undefined commands are ignored 
and no reply is sent.