10. COMPARATOR 
 
Five-level and three-level comparators are available. 
The five-level comparator mode uses four comparator values to compare the weight 
value and outputs results in five levels of LOLO, LO, OK, HI and HIHI. 
The three-level comparator mode uses two comparator values (upper and lower limit 
values) to compare the weight value and outputs results in three levels of LO, OK and HI. 
The results are output using LEDs (yellow / green / red). 
When the optional comparator relay output (OP-03) is installed, the results are output as 
a relay signal. 
 
‰
 To use the comparator modes, the function settings “
Cp-l
” and “
Cp
” must be 
specified. 
‰
 Using the function setting “
Cp-l
”, select a comparator mode. 
0: five-level comparator mode 
1: three-level comparator mode 
‰
 Using the function setting “
Cp
”, select comparison conditions.
 
0: No comparison (comparator mode disabled). 
1: To compare all data. 
2: To compare all stable data. 
3: To compare all data which are more than or equal to +5d, or less than 
or equal to -5d. 
4: To compare stable data which are more than or equal to +5d, or less 
than or equal to -5d. 
5: To compare all data which are more than or equal to +5d. 
6: To compare stable data which are more than or equal to +5d. 
d = minimum display in kg (Refer to “15.1. Specifications”.) 
Also in the counting mode, “d” is equal to the minimum display of kg mode. 
 
10.1. The formula to compare   
Comparison is performed using the formula listed below and the results are output. 
Five-level comparator mode 
Results 
Comparison formula 
LED display 
LOLO 
Displayed value < LOLO limit value   
(Or over in the negative value) 
 
( Red LED on) 
LO 
Displayed value < LO limit value   
 
( Yellow LED on) 
OK 
LO limit value ≤ Displayed value ≤ HI limit value   
 
( Green LED on))
HI 
HI limit value < Displayed value 
 
( Yellow LED on) 
HIHI 
HIHI limit value < Displayed value 
(Or over in the positive value) 
 
( Red LED on) 
 
LOLO limit value = Secondary lower limit value 
LO limit value = Lower limit value 
HI limit value = Upper limit value 
HIHI limit value = Secondary upper limit value