‰ The HC-i is designated as DCE (Data Communication Equipment). 
13-3. Data Output Mode 
‰ The Data Output Modes and Parameters are set by F-Functions in F-06-X as 
described in the “9-2. F-Functions”
‰ To control the scale using commands form an external device, see “13-6. Command 
Mode”.
‰ Refer to “13-6. Command Mode” about the output data format. 
Data Output Mode (f-06-01)
‰ Key Mode (f-06-01="0")
When the weight display is stable, data is sent by pressing the  PRINT   key. The 
count display will blink when the data has been sent. 
‰ Stream Mode (f-06-01="1") 
Data is sent continuously. The data-update rate is approximately 10 times per second for 
f-06-03="2". For f-06-3="0" or "1", the interval between continuous data is 
approximately 2 seconds. 
‰ Auto-print Mode A (f-06-01="2")
Data is sent if the weight display is stable at +5d (weighing display division) and 
above. The next transmission can not occur until after the weight display falls below 
+5d.
‰ Auto-print Mode B (f-06-01="3")
Data is sent if the weight display is stable at ±5d (weighing display division) and 
above/below. The next transmission can not occur until after the weight display falls 
between –5d and +5d. 
To use with the UFC format, refer to "13-7. Using UFC (Universal Flex Coms) 
Function"
Data to be Sent (f-06-02)
Select which data to be sent by keying in a  0  or  1  for the data: ID No., PCS 
(count), weight or unit weight. 
Example: Key in  1
1
0
0  to 
display 1100, this setting would 
send only the ID number and the 
count.
Data Format (f-06-03)
‰ Format for AD-8121 MODE 1 or 2. (f-06-03="0") 
‰ Format for AD-8121 MODE 3. (f-06-03="1")
‰ Format for general apparatuses, computers, etc. (f-06-03="2") 
‘0’ not to send data
‘1’ to send data 
8
f-06-02
88Clear8
8
ai
1100
0
00
ID No.
PCS (count)
subtr