A&D AD-4405 Benutzerhandbuch

Seite von 80
 Page 
12 
AD-4405 Weighing Indicator
 
Note that the 
ON/OFF
 key does not function alone. Press the 
ESC
 key while 
holding the 
ON/OFF
 key to end the calibration mode, if mis-operation. After 
displaying  Can5el, press the 
ON/OFF
 key to end the calibration mode and 
turn the indicator off. 
NOTE: When 
displaying 
Caloff , press the +/- key while pressing the 
ON/OFF
 key, 
instead of 
CAL
 key. However, in some cases, settings are altered to nullify 
this option. 
 
 
Caution 
  The maximum display is less than or equal to 40000 divisions. This number 
is calculated from the maximum capacity divided by the minimum division. 
 
  Check the accuracy of the weighing instrument periodically.   
 
  Recommended mass, use a mass heavier than 2/3 maximum capacity. 
 
  Calibrate the scale, if it is moved to another location or the environment 
has changed. 
 
  It is not necessary to set the gravity acceleration correction, when calibrating 
the scale with a calibration mass at the place where the scale is used. 
 
  Enter the stable weighing data while the STABLE mark is displayed. If 
unstable data is used, it may cause a weighing error. Arrange the 
condition using the F00 filter function. 
 
  The span calibration needs the zero calibration data. We recommend 
that you perform the span calibration immediately after the zero 
calibration. 
 
  If you use the dual range function of the multi-interval scale, perform the 
"Range Function", "Zero Calibration" and "Span Calibration". 
 
 
5.2.
5.2.
5.2.
5.2.    Calibration Procedure
Calibration Procedure
Calibration Procedure
Calibration Procedure    
5.2.1. 
Configuring a Weighing Instrument 
 This section explains how to set capacity, resolution, decimal point position and display 
format, weighing range and unit. Perform this procedure when installing the indicator. 
 
 When  Cal5et appears on the display, enter the setting mode by pressing of the 
ENTER key. 
Setting the range and unit. 
 
Single Range 
Select the resolution, decimal point position and format. 
¯ 
Specify the weighing capacity. 
 
5. Calibration