A&D AD-4405 Benutzerhandbuch

Seite von 80
 Page 
32 
AD-4405 Weighing Indicator
 
 
6.3.
6.3.
6.3.
6.3.    CF
CF
CF
CF-
-
-
-Functions
Functions
Functions
Functions    
 
Item 
Parameter 
Description 
* 0 
No limitation 
Cf00 
Zero tracking width, 
motion detection 
condition 
Use limitation at F01, F02, F03, F27 and F28. 
* 0 
2% of CAP, 
Tare limit is 100 % CAP 
10% of CAP,   Tare limit is 100 % CAP 
3% of CAP, 
Tare limit is 50 % CAP 
Cf01 
Zero range 
Turning display on, the 
range to zero display. 
4% of CAP, 
Tare limit is 50 % CAP 
Not to zero when turning the display on. 
* 1 
10% of CAP   
  3% of CAP 
Cf02 
Power on zero range 
Turning the display on, 
the range to zero display. 
  4% of CAP   
Gross when displaying gross. 
Gross 
Cf03 
Zero tracking 
* 2 
Gross or Net when displaying net. 
 
TARE, ZERO in motion / TARE at negative gross 
* 0 
Not accepted 
Not accepted 
Execute / 
Not 
accepted 
Not accepted 
Execute 
Cf04 
TARE, ZERO in motion / 
TARE at negative gross 
value 
Execute / 
Execute 
* 0 
Not to output data at unstable value or overload.   
Effective in key mode. 
Cf05 
Output on overload and 
unstable state. 
To output data always.   
 
GROSS  / NET  / TARE / Preset TARE 
* 0 
GR  / NT / TR  / 
TR 
GR  / NT / TR  / 
PT 
Cf06 
Header 2
 
G_ / 
N_ / T_ 
PT        (_: Space 20h)
* 0 
Two digits 
Cf07 
Number of unit digits 
Three digits 
Serial output data 
* 0 
Not used (Ineffective) 
Cf08 
Accumulation function 
Used    (Effective) 
* 0 
Used    (Effective) 
Cf09 
Digital Tare(PT) 
Not used (Ineffective) 
 
CAP : maximum capacity 
*: Initial settings 
 
Power on ZERO (CF02) 
Cf02 0 
After power on, the weight display starts immediately. 
Cf02 1,2,3 
When the zero range is exceeded at power on,   ------  is displayed. 
 Press 
the 
ESC
 key to start the weight display. 
 
6.3. CF-Functions