A&D AD-4405 Benutzerhandbuch

Seite von 80
AD-4405 Weighing Indicator
 Page 
47 
 
 
ENTER
 key..............The key to register the unit weight and return to the piece 
count display. The unit weight is calculated automatically 
with the weight and the number of samples. 
 
 
 Press 
the 
ESC
 key while pressing 
+/-
 key 
   
 
The key to return to the piece count display without changing 
the unit weight. 
Notes 
  It is preferable to have a large number of samples, to minimize the 
counting error. 
 
 Press the 
ENTER
 key after the STABLE mark turns on. 
 
  If the total weight of the samples is too light and is not adequate to be 
used as the unit weight, 
    loUt
loUt
loUt
loUt      is displayed for 2 seconds and returns 
to the registration display. 
 
Digital input 
 
When the unit weight is already known, digital input is available.   
Step  1  Start with the normal weight mode. 
Step 2  Press the 
SAMPLE
 key to enter the digital input of the unit the weight. U xxx and 
the weight unit is displayed with unit 
pcs
pcs
pcs
pcs blinking. Using following keys: 
 
0
 - 
9
 key..................The key to set the unit weight. Maximum is 5 digits. 
 
+/-
 key .....................The key to select the weighing unit of the unit weight. 
 
ZERO
 key................The key to clear the digital input. U---- is displayed. 
 
ESC
 key ..................The key to return to the normal weight display without 
changing the registered unit weight. 
 
ENTER
 key..............The key to register the unit weight and return to the normal 
weight display. 
 Press 
the 
MODE
 key while pressing the 
+/-
 key The key to shift the decimal point. 
 
 
 
12. Counting Function