Tanita BD-815U Benutzerhandbuch

Seite von 15
GB
GB
7
6
Safety Notes 
(continued)
To measure correctly
• Do not use cellular phones or microwave therapy equipment near this equipment.
It may not measure correctly due to equipment malfunction.
Do not measure at a place where the temperature fluctuates
greatly.
Errors may be observed in the measured value.  In the case to move to a
place with a temperature difference of 36˚F (20˚C) or more, leave it
before use for 2 hours or more after moving.
Please observe the following instructions for accurate measurement.
• Measurement may not be possible on a surface that is strongly vibrating. In such cases
transfer the equipment to a stable surface with little or no vibration.
• In the case to move to a place with a temperature difference of 36˚F (20˚C) or more, leave
it before use for 2 hours or more after moving.
• While taking measurements, please keep any persons using transmitters, such as a 
cellular phone, away from the unit to avoid causing margin errors.
<Usage Conditions>
Temperature Range for Use : 41˚F - 95˚F (5˚C - 35˚C)
Relative Humidity
: 30% - 80% (without condensation)
<Storage Conditions>
Temperature Range of Environment : 14˚F - 122˚F (-10˚C - +50˚C)
Range of Relative Humidity
: 10% - 90% (without condensation)
To avoid malfunctions, avoid storing the equipment where there is direct sunlight,
significant temperature changes, the risk of dampness, a large amount of dust, in
the vicinity of fires, or where there is the risk of receiving vibrations or shocks.
Instructions for Measurement
Explanation of terminology
• Capacity
This shows the maximum weight that can be measured by the scale.
[Example]
In the case of Maximum capacity 33lb (15kg), it can measure up to 33lb (15kg).
* One exceeding the weighing capacity cannot be measured.
• d
This shows the amount of one scale mark.  In the case of a digital display, this shows the inter-
val of the numeric value to change.
[Example]
In the case of Minimum Graduation 0.1oz (0 - 12lb) 0.2oz (12 - 33lb) or 2g (0 - 6kg) 5g (6 - 15kg)
the display increases and decreases.
• Extend Minimum Graduation
Enables changing precision of measurements.
• Tare
The container, towel, blanket, etc., used when measuring weight is called “tare”.
To measure the net-weight after subtracting the weight of the tare is called "tare weighing."
<AC Adapter>
Model
AD-0920M
Input
120 VAC
60Hz
Output
9 VDC
200mA