6. COMPARATOR 
 
The results of the comparison are indicated by HI, OK or LO on the display. 
The comparison is as follows: 
 LO 
< Lower limit value ≤ OK ≤ Upper limit value < HI 
 
Operating conditions (see the function setting “Cp”): 
‰
  No comparison (comparator function disabled). 
‰
  Compares all data. 
‰
  Compares all stable data. 
‰
  Compares plus data except those near zero (plus data greater than +4d). 
‰
  Compares stable plus data except those near zero (stable plus data 
greater than +4d). 
‰
  Compares all data except those near zero (all data greater than +4d or 
less than -4d). 
‰
  Compares stable data except those near zero (stable data greater than 
+4d or less than -4d). 
d = the smallest display division.    e.g.: 4d = four display divisions 
 
The upper limit and lower limit numerical values are common to each of the weighing, 
counting and percent mode. The example for EJ-120/200/300/410/610 is as follows. 
 
Upper limit value “001010”:  “10.10g” “1010pcs”  “101.0%” 
Lower limit value “000990”:   “9.90g”   “990pcs”   “99.0%” 
 
 
6-1. Setting example 
 
This example will be “Compares plus data except 
those near zero”. 
 
Selecting a comparison mode   
 
1. Press and hold the    SAMPLE    key to display 
  func      . 
(If the comparison mode is already set, press the 
 SAMPLE  key to go to “Entering the upper and 
lower limit values”.) 
 
2. Press the    PRINT    key, then the balance displays 
 poff X . 
 
3. Press the    SAMPLE    key several times to display 
  Cp   X . 
 
4. Press the    RE-ZERO    key several times to display 
 Cp   3 . 
 
5. Press the    PRINT    key to store the settings. 
 Cp Hi   appears after  end    . 
 
 
 poff 0 
 Cp   0 
 Cp   3 
Press and hold
Press
several times
Press
several times
To store
 func