A&D EJ-1500 Benutzerhandbuch

Seite von 42
 25
 
‰
 Key mode   
Set the function “prt 1, 2 or 3”. 
When the  PRINT  key is pressed while the weighing data is stable (the STABLE  
indicator is on), the balance transmits the data. When the data is transmitted, the 
display will blink one time. 
 
‰
 Auto-print mode A 
Set the function “prt 2”. 
The balance transmits the weighing data when the display is stable (the STABLE 
indicator is on) and the data is greater than +4d. 
The next output can be obtained after the display returns below +4d. 
 
‰
 Auto-print mode B 
Set the function “prt 3”. 
The balance transmits the weighing data when the display is stable (the STABLE 
indicator is on) and the data is greater than +4d or less than -4d. 
The next output can be obtained after the display returns between -4d and +4d. 
 
 
Command mode   
In the command mode, the balance is controlled by commands that come from the 
personal computer and so on. 
 
Command list   
‰
  Command to request the current weighing data. 
Command Q 
C
R
 L
F
 
 
Reply  S 
+ 0 0 1 2 3 4 . 5
  g C
R
 L
F
 
 
‰
  Command to zero or tare the balance (same as the    RE-ZERO    key). 
Command Z 
C
R
 L
F
 
 
Reply Z 
C
R
 L
F
 
 
‰
  Command to change the weighing units (same as the    MODE    key). 
Command U 
C
R
 L
F
 
 
Reply U 
C
R
 L
F