A&D EJ-1500 Benutzerhandbuch

Seite von 42
 31
10-2. Output example 
‰
 To print the GLP report, set the function “info 1” and use MODE 3 of the printer 
AD-8121B. 
‰
 To output the GLP report to a personal computer, set the function “info 2”. 
 
Data format for “calibration report” 
Perform calibration using a weight. Then the balance will output a calibration report. 
See “7-1. Calibration using a weight” about the calibration. 
 
 AD-8121 format “info 1” 
General format “info 2” 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
Data format for “calibration test report” 
Perform calibration test using a weight. Then the balance will output a calibration test 
report. See “10-3. Calibration test using a weight
 
” about the calibration test. 
 
 AD-8121 format “info 1” 
General format “info 2” 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
  :Space, 
ASCII 
20h 
 
CR :Carriage return, ASCII 0Dh 
 
LF  :Line feed, ASCII 04Ah 
 
Manufacturer
Model
Serial number
ID number
Date
Time
Calibration
executed 
Calibration weight
Signature
           A & D 
MODEL     EJ-300 
S/N      1234567 
ID        ABCDEF 
DATE    09/03/21 
02:53:21 PM 
CALIBRATED(EXT.) 
CAL.WEIGHT 
      +300.00  g 
SIGNATURE 
 
 
- - - - - - - -  
~~~~~~~~~~~
A
~
&
~
D<CRLF>
MODEL
~~~~~
EJ-300<CRLF>
S/N
~~~~~~
1234567<CRLF>
ID
~~~~~~~~
ABCDEF<CRLF>
DATE<CRLF> 
<CRLF> 
TIME<CRLF> 
<CRLF> 
CALIBRATED(EXT.)<CRLF>
CAL.WEIGHT<CRLF> 
~~~~~~
+300.00
~~
g<CRLF>
SIGNATURE<CRLF> 
<CRLF> 
<CRLF> 
-
~
-
~
-
~
-
~
-
~
-
~
-
~
-
~
<CRLF>
<CRLF> 
<CRLF> 
           A & D 
MODEL     EJ-300 
S/N      1234567 
ID        ABCDEF 
DATE    09/03/21 
03:15:40 PM 
CALIBRATED(EXT.) 
ACTUAL 
         0.00  g 
      +299.99  g 
TARGET 
      +300.00  g 
SIGNATURE 
 
 
- - - - - - - -  
~~~~~~~~~~~
A
~
&
~
D<CRLF>
MODEL
~~~~~
EJ-300<CRLF>
S/N
~~~~~~
1234567<CRLF>
ID
~~~~~~~~
ABCDEF<CRLF>
DATE<CRLF> 
<CRLF> 
TIME<CRLF> 
<CRLF> 
CAL.TEST(EXT.)<CRLF> 
ACTUAL<CRLF> 
~~~~~~~~~
0.00
~~
g<CRLF>
~~~~~~
+229.99
~~
g<CRLF>
TARGET<CRLF> 
~~~~~~
+300.00
~~
g<CRLF>
SIGNATURE<CRLF> 
<CRLF> 
<CRLF> 
-
~
-
~
-
~
-
~
-
~
-
~
-
~
-
~
<CRLF>
<CRLF> 
<CRLF> 
Manufacturer
Model
Serial number
ID number
Date
Time
Calibration
test
Signature
Zero value
Weight value
Target weight value