MODEL 315  HUMIDITY TEMPERATURE METER 
 
4.0 Operating Instructions: 
4.1 Preparation 
for 
Measurement: 
To measure ambient humidity and temperature, hold the instrument in the 
environment of interest and allow sufficient time for the sensors to respond 
to ambient conditions. (See Section 2.0 – Response Time.) 
4.2 Power: 
Press the        button to turn the instrument ON or OFF. 
4.3 Selecting the Temperature Scale 
When first turned on, the instrument defaults to reading in Celsius (°C). The 
instrument toggles between Celsius (°C) and Fahrenheit (°F) each time the 
°C/°F button is pressed. The instrument remembers the scale setting when 
last turned off and powers on in that setting the next time. 
4.4 Data-Hold 
Function 
The present reading is held on the display by pressing the HOLD button. 
Pressing the HOLD button again releases the hold function and returns the 
instrument to continuous reading.   
In the HOLD mode, the 
UREL, MAX MIN and °C/°F buttons are disabled, 
as indicated by two consecutive beeps when pressed. 
4.5 Relative 
Operation 
 
Pressing the 
UREL button internally memorizes
 
the present
 
reading and 
the main display shows the difference between the new reading and the 
memorized value.  Press the 
UREL button again to exit the Relative 
operation. 
4.6 MAX/MIN 
Operation 
Press the MAX MIN button to enter the MAX/MIN mode. In this mode, both 
the maximum and minimum values are simultaneously retained in memory 
and updated with every new data sample. The instrument first enters the 
MAX mode, and the MAX symbol and value are both displayed. Pressing 
MAX MIN again advances the display to the MIN symbol and value. The 
next press of the MAX MIN button will cause both the MAX and MIN 
symbols to blink. This indicates that the maximum and minimum values 
have been updated in memory and the displayed reading is the present 
value. Each successive
 
press of the MAX MIN button circulates the display 
mode among these options. To exit the MAX/MIN mode, press and hold 
MAX MIN for two seconds. 
In the MAX/MIN mode, the
UREL  and  °C/°F buttons are disabled, as 
indicated by two consecutive beeps if pressed. 
 
MODEL 315  HUMIDITY TEMPERATURE METER 
 
4.7 Auto 
Power 
Off 
By default, the instrument powers on in the ‘Auto Power Off’ mode and will 
automatically shut off 30 minutes after the last key operation. 
To disable this feature, press and hold the HOLD button and then power on 
the meter. Two successive beeps indicate that ‘Auto Power Off’ is disabled. 
4.8  Low Battery Condition 
When the battery voltage is at or below the minimum for proper operation, 
the        symbol will show on the display indicating that the battery must be 
replaced. 
4.9 Calibration 
Points: 
Room temperature: 23°C ± 5°C 
Temperature calibration points: 0°C and 40°C 
Humidity calibration points: 32.8% RH@25°C and 75.3% RH@25°C 
 
 
5.0 Calibration Procedure: 
Note: Conduct temperature calibration before humidity calibration. 
5.1 Temperature 
Calibration 
1.  Turn the unit off. Press and hold the HOLD and °C/°F buttons. While 
these two buttons are pressed, turn the power on. Release all buttons 
and all the segments on the display will blink. 
2.  Within 3 seconds of releasing these buttons, again press the °C/°F 
button to enter the calibration mode, or the instrument will revert to 
normal operating mode. 
3.  Upon entering the calibration mode, ‘CAL1’ will appear on the second 
display.  
4.  Insert the probe into a temperature-controlled chamber at 0°C (32°F) 
and allow the system to stabilize for 10 minutes. Press the MAX MIN 
button to initiate calibration. A single long beep will indicate calibration 
is complete, and the instrument will advance into the ‘CAL2’ mode, as 
indicated on the second display. 
5.  Insert the probe into a temperature-controlled chamber at 40°C (104°F) 
and again allow the system to stabilize for 10 minutes. Press the   
MAX MIN button to initiate calibration. A single long beep will indicate 
calibration is complete, and the instrument will revert to normal 
operating mode.