Tektronix 070-9472-02 Benutzerhandbuch

Seite von 56
Specifications
P5205 Instruction Manual
17
Table 3: Certifications and Compliances (Cont.)
Installation Category
Descriptions
Terminals on this product may have different installation category
designations. The installation categories are:
CAT III
Distribution-level mains (usually permanently
connected). Equipment at this level is typically
in a fixed industrial location
CAT II
Local-level mains (wall sockets). Equipment at
this level includes appliances, portable tools,
and similar products. Equipment is usually
cord-connected
CAT I
Secondary (signal level) or battery operated
circuits of electronic equipment
Pollution Degree 2
Do not operate in environments where conductive pollutants may
be present.
Typical Characteristics
Typical characteristics in Tables 4 and 5 describe typical but not
guaranteed performance.
Table 4: Typical Electrical Characteristics
Rise Time
3.5 ns
Bandwidth Limit
5 MHz
AC Common-Mode Rejection Ratio
(20
-
30
_C, <70% RH)
See Figure 5 on page 18
60 Hz: > 10,000:1
100 kHz: > 300:1
100 MHz: > 300:1
AC Noise
(referenced to input)
50X: < 50 mV
RMS
500X: < 300 mV
RMS
Input Impedance
8 M
Ω, 3.5 pF between inputs
4 M
Ω, 7 pF between each input and ground
See Figure 6
Propagation Delay
17 nS