Safety Information
The 80T-IR/E complies with IEC Publication 1010-1-
1990 including Amendment 1, CSA C22.2 No. 231,
ANSI/ISA-S82.01 and .03 Safety Standards.
1990 including Amendment 1, CSA C22.2 No. 231,
ANSI/ISA-S82.01 and .03 Safety Standards.
W
WARNING
IF TARGET EMMISSIVITY IS LESS THAN 0.95, THE
PROBE CAN INDICATE A TEMPERATURE LOWER
THAN THE ACTUAL TARGET TEMPERATURE.
AVOID TOUCHING THE TARGET; THERMAL BURNS
COULD RESULT.
PROBE CAN INDICATE A TEMPERATURE LOWER
THAN THE ACTUAL TARGET TEMPERATURE.
AVOID TOUCHING THE TARGET; THERMAL BURNS
COULD RESULT.
CAUTION
·
Do not place the probe on or around hot objects (70°C
/ 158°F). It will damage the probe case.
·
If the probe is exposed to significant changes in
ambient temperature (hot to cold or cold to hot),
allow 20 minutes for temperature stabilization,
before taking measurements.
allow 20 minutes for temperature stabilization,
before taking measurements.
·
Do not operate the probe near large electrical or
magnetic fields such as arc welders and
induction heaters. These fields can cause
measurement errors.
induction heaters. These fields can cause
measurement errors.
·
Condensation may form on the lens when going
from a cold to hot environment - wait 10 minutes
for condensation to dissipate before taking
measurements.
for condensation to dissipate before taking
measurements.
·
Connectors must only be plugged into voltage
measurement input jacks of the test instruments.
·
Do not touch or hold by the front cone.
Temperature readings can be affected by heat
from the hand.
from the hand.
·
Equipment use not specified by manufacturer
may impair safety.