Displays error message: 
SYSTEM ERRxxxx Call Service!
 or Exxxx (where 
xxxx defines an error)
Displays error message: 
Check tube!! Key if done
The following message is displayed on the LCD after you try to update settings 
or copy data to the Data Card: 
Card Error Remove Card
The following message is displayed on the LCD after you try to update settings 
or copy data to the Data Card: 
Settings Error Remove Card
The following message is NOT displayed on the LCD after you try to update the 
settings using the Data Card: Settings Success Remove Card
The following message is displayed on the LCD after you try to update settings 
or copy data to the Data Card: 
Settings Invalid Remove Card
Problem/Possible Cause
Solution
Component failure.
Record error number and call service.
Air tubing is loose or blocked.
Check that the air tubing is connected securely to your 
mask and the air outlet to the front of the S8 Series II. 
When you have checked the air tubing, press the Start/
Stop key to restart the device. If this does not clear the 
message, disconnect the power cord and then 
reconnect it to restart the device.
Data Card is not inserted 
correctly.
Check that the patient is inserting the Data Card into the 
Data Card module with
 
the arrow facing up, while the S8 
Escape II is switched on and not delivering therapy 
(standby mode).
The Data Card is only partly 
inserted.
Ensure that the Data Card is inserted into the Data Card 
slot as far as it will go.
You may have removed the 
Data Card before settings 
were copied to the 
S8 Escape II.
Reinsert the Data Card and wait for the Settings 
Success Remove Card
 message to appear on the 
LCD.
The card is not a ResScan Data 
Card.
Only use a ResScan Data Card.
There is a data error on the 
Data Card.
Contact your clinician immediately.
The settings were not 
updated.
Contact your clinician immediately.
The identification details on 
the Data Card do not match 
the identification details on 
your device.
Contact your clinician immediately.