3M 2273M Benutzerhandbuch

Seite von 66
78-8130-6150-0 Rev G 
23
B. Dyna-Coupler Method
! WARNING
This WARNING applies to the following 3M
 Dyna-Couplers;
• 3" (75 mm) - Part number 3001
• 4.5" (114 mm) - Part number 4001
• 6" (150 mm) - Part number 1196
•  All accessory kits containing any of the listed 3M Dyna-Couplers - Part numbers 3019, 4519, 
1196/C
A potential for electrical shock exists when using the 3M Dyna-Coupler on cables energized 
with electrical power. Use appropriate safety procedures.
DO NOT USE ON CABLES CARRYING IN EXCESS OF 600 VOLTS RMS.
Use a ground extension cable, 
or wire, and ground rod set-up 
between a target pipe/cable and 
a gas meter valve/box, or cable/
electrical meter box. This will 
isolate the locate signal to the 
target pipe/cable section below 
the meter/box, between the 
grounding points. This provides a 
good return path for the tracing 
signal. Insulating coupling above 
a gas meter valve on a pipe will 
isolate the returning signal from 
ground and may make locating 
more difficult.
1.  Connect the 3M Dyna-Coupler to the transmitter Output Jack [T-6] using the 
coupler cable (9011).
2.  Clamp the 3M Dyna-Coupler around the cable or pipe, below any bonds, just before 
it enters the earth. The jaws of the coupler must fully close.
3.  Press Trace [T-3] to turn on the transmitter. Press again to select 8 kHz, 33 kHz or 
200 kHz.
Note: When using a 3M Dyna-Coupler, always select high or maximum output power 
level by pressing the Output Level [T-5] key on the transmitter.