WARNING: 
This guide is only a summary of the steps necessary to properly disinfect your endoscope. Be sure to follow the 
detailed steps outlined in the ENDOSCOPE REPROCESSING MANUAL that was included with your endoscope when purchased. 
To use this chart as a wallchart, cut  
off the top of the chart as illustrated.
To insert this chart into a binder, punch out 
the holes and fold the chart as illustrated.
Suction Cleaning Adapter
Tube to suction
To Suction Pump
Suction Cleaning
Adapter (MAJ-222)
To Suction Pump
Clean the bronchoscope 
meticulously prior to disinfection.
Single-Use Suction Valve (MAJ-209)
Suction Cylinder
Single-Use Biopsy Valve (MAJ-210)
Remote Switches
Universal Cord
Instrument/Suction Channel Port
Electrical Connector
Water-Resistant Cap 
(MH-553)
Venting Connector
Distal Tip
Bending Section
Light Guide
Boot
UP/DOWN Angulation Control Lever
Control Section
Insertion Tube
Alcohol
HIGH-LEVEL DISINFECTION
RINSING
TRANSPORTATION
DRYING
NOMENCLATURE
IMPORTANT
1
2
3
4
5
6
9
10
7
8
©
 2
0
0
4
 O
ly
m
pu
A
m
er
ic
In
c.
 /
 P
ri
nt
ed
 in
 U
SA
  
0
1
/0
4
 –
 P
W
C
-1
0
0
0
1
6
EXERA
 BF-160 Series and EVIS
 BF-240 Series Bronchovideoscopes
BRONCH
OSC
O
PE DISINFECTION GUIDE
 Refer to the companion BRONCHOSCOPE CLEANING GUIDE for detailed cleaning information
WARNING:
 
This guide is only a summary of the steps necessary to properly disinfect your bronchoscope. Be sure to follow the 
detailed steps outlined in the ENDOSCOPE REPROCESSING MANUAL that was included with your bronchoscope when purchased. 
Attach a syringe to the adapter. 
Pull the plunger of the syringe to
fill the instrument/suction channels
and channel cleaning adapter
with disinfectant solution.
Fill basin with disinfectant.  Attach a 
suction cleaning adapter (MAJ-222) 
to the previously cleaned broncho-
scope.  Immerse completely in 
disinfectant solution.
While immersed, detach the adapter 
from the bronchoscope. Use a clean, 
lint-free cloth to wipe away any bub-
bles clinging to external surfaces.
Soak all equipment according to
the time and temperature 
recommended by the disinfectant 
manufacturer. Use a timer to 
measure the exact contact time.
Connect the suction cleaning
adapter and syringe. Inject air 
to remove disinfectant from the 
instrument/suction channels, 
then remove the bronchoscope 
from the disinfectant. 
Disconnect the suction cleaning
adapter and immerse the 
bronchoscope and adapter in 
sterile water. Using a sterile, 
lint-free cloth, thoroughly rinse 
and wipe all surfaces.
Connect suction cleaning adapter 
to bronchoscope and to a tube 
from a suction pump. Aspirate 
water through channels for 30 
seconds. Remove from water and 
aspirate air through channels for 
60 seconds.
Turn the suction pump off. Hold the control 
section with the openings pointing down and 
disconnect the suction cleaning adapter.
Dry the bronchoscope and adapter with an 
alcohol-moistened cloth. Make sure that the 
equipment is completely dry prior to storage.
Hold all parts of the endoscope when carrying. Do 
not let the ends of the bronchoscope swing freely.
• Wear appropriate personal protective 
equipment during reprocessing.
• Autoclave all autoclavable accessories.
• Store the bronchoscope hanging vertically 
with all valves and caps removed.
• Use a compatible disinfectant and test
it for proper concentration as recom-
mended by the manufacturer.
• Do not use the carrying case to store 
"patient-ready" endoscopes. Any broncho-
scope placed in the carrying case must be 
reprocessed prior to patient use.
Immerse the bronchoscope's distal 
end in 70% alcohol and aspirate 
alcohol for 5 seconds. Remove the 
distal end from the alcohol and 
aspirate air for 20 seconds.