Calibration Procedure
 
05206H DCN5910
 
17
 
 
 
Table 2-2. Zero Calibration Procedure - Zero Gas Through ZERO Port 
 
Step No. 
Action 
Comment 
1. Press 
CALZ 
The M201E enters the calibrate mode from sample 
mode. The zero gas is supplied through the ZERO
inlet port on the rear panel of the M501 NH
3
,. 
Table 2.1 
 
Notes: 
  Since the zero gas concentration is defined as 0 ppb, it is not necessary to 
enter the expected concentration values. 
  Both the TN
x
 and NO
x
 channels must be zeroed individually. When you 
zero the NO
x
 channel both the offsets for NO
x
 and NO are configured. All 
readings should display zero following this action. 
  It is recommended that the calibration of both the TN
and NO
x
 channels 
be done at one time.  
  Always confirm that excess zero air is flowing past the sample or zero 
port, if the Zero/Span option is present, on the rear of the M501 NH
3
converter. There should always be some excess flow. If insufficient flow is 
supplied to the analyzer ambient gas will be entrained and the zero offset 
values will be incorrect. However, it is important not to over pressurize the 
analyzer. 
  It is a good idea to allow the analyzer to sample zero gas for extended 
periods following zero calibration. Large drifts can signify leaks or indicate 
an issue with the zero air source. 
 
2.2  S
PANNING 
T
HE 
A
NALYZER 
W
ITH 
N
ITRIC 
O
XIDE 
G
AS
 
The analyzer can be spanned by either applying nitric oxide gas straight to the 
Sample port or through the Zero/Span valve option, if installed. It is important to 
remember that if the analyzer was previously sampling
 
ammonia gas, prior to 
conducting a nitric oxide span calibration, it will take some period of time before 
the TN
x
 channel stabilizes. The process of spanning the analyzer consists of 
diluting bottled gas, using a calibrator and delivering to the analyzer. The