1-2       Description and Specifications                                                Rosemount Analytical Inc.    A Division of Emerson Process Management 
Model NGA2000 PMD
1-4 FEATURES 
Among the features incorporated into the 
PMD Analyzer Module is a flow splitter (
≈2:1) 
that allows for greater sample flow, decreased 
lag time and faster analyzer response. 
The "Time Alignment" feature can be used to 
delay sending the Primary Variable from the 
PMD Analyzer Module for up to 30 seconds in 
0.1 second intervals.  This feature allows Pri-
mary Variables form more than one PMD 
Analyzer Module to be "time aligned" if nec-
essary.
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
Figure 1-2.    Trace Oxygen Detector Coulometric Principle 
SIGNAL CONDITIONING
ASSEMBLY
Dual Photocell
Displacement
Torque
Restoring
Torque
Magnet
Platinum/Nickel Alloy
Suspension Ribbon
Balancing
Weight
Balancing Weight
Nitrogen-Filled Hollow Glass
Test Body
Mirror
Current Loop
TEST BODY DETAIL
Electromagnetic
Axis
Electromagnetic
Axis
Restoring
Current
Restoring
Current
Displacement
Torque
Restoring
Torque
DETECTOR/MAGNET
ASSEMBLY
Test Body
Shaded Pole Pieces (4)
LED Source
Restoring
Current
Span
Zero
SIGNAL OUTPUT
TO NETWORK
NULL BALANCE SYSTEM