Emerson NGA2000 PMD Benutzerhandbuch

Seite von 88
Instruction Manual 
760007-A 
July 2003 
3-2     Operation 
Rosemount Analytical Inc.    A Division of Emerson Process Management 
Model NGA2000 PMD
 
 
 
 
 
 
 
MESSAGE DISPLAY 
DESCRIPTION 
TYPE 
barometer System 
Barometer 
warning 
case temp 
Case Temperature 
warning 
crude noise 
Calculated Noise 
warning 
currentrnghi 
Current, High Range 
warning 
currentrnglo Current, 
Low 
Range 
warning 
det tem 
Detector Temperature 
warning 
fan fet 
Fan Current 
warning 
heater fet 
Heater Current 
warning 
led current 
LED Current 
warning 
lin error 
Linearizer Error 
warning 
loop current 
PMD Loop Current 
warning 
n15 volts 
Power Supply, -15V 
warning 
p15 volts 
Power Supply, +15V 
warning 
p24 volts 
Power Supply, +24V 
warning 
P5 volts 
Power Supply, +5v 
warning 
raw signal 
Raw Signal 
warning 
Samp Pres 
Sample Pressure 
warning 
svflow 
Sample Bypass Flow 
warning 
bicella 
PMD Photo Sensor 
failure 
bicellb 
PMD Photo Sensor 
failure 
sw error 
Software Error 
failure 
 
 
Table 3-1.  PMD Analyzer Module Alarms