ResMed S8 ESCAPE SYSTEM Benutzerhandbuch

Seite von 39
23
E
NGLISH
C
LEANING
 
AND
 M
AINTENANCE
R
EPLACING
 
THE
 A
IR
 F
ILTER
To replace the air filter:
1. Remove the air filter cover at the back of the 
S8 E
SCAPE
.
2. Remove and discard the old air filter.
3. Insert a new filter with the blue tinted side facing out from the device.
4. Replace the air filter cover.
Inspect the air filter every month to check if it is blocked by dirt or contains holes. With 
normal use of an 
S8 E
SCAPE
, the air filter needs to be replaced every six months (or 
more often if your device is in a dusty environment).
!
WARNING
Do not wash the air filter. The air filter is not washable or reusable.
The air filter cover protects the device in the event of accidental liquid spillage 
onto the device. Ensure the air filter and air filter cover are fitted at all times.
S
ERVICING
!
CAUTION
Do not attempt to open the S8 Escape case. There are no user serviceable 
parts inside. 
Inspection and repair should only be performed by an authorized agent. Under 
no circumstances should you attempt to service or repair the flow generator 
yourself.
This product (
S8 E
SCAPE
) should be inspected by an authorized ResMed Service 
Center five years from the date of manufacture. Prior to this, the device is intended to 
provide safe and reliable operation provided that it is operated and maintained in 
accordance with the instructions provided by ResMed. Applicable ResMed warranty 
details are provided with the device at the time of original supply. Of course, as with 
all electrical devices, if any irregularity becomes apparent, you should exercise caution 
and have the device inspected by an Authorized ResMed Service Center.
If you feel that your flow generator is not performing properly, see “Troubleshooting” 
on page 26.