If oxygen is used with this device, the oxygen flow must be turned off when the 
device is not operating.
If oxygen has been left on, turn off the flow generator, then wait 30 minutes before 
turning on the flow generatory again.
Explanation: When the CPAP device is not in operation and the oxygen flow is 
left on, oxygen delivered into the air delivery tubing may accumulate within the 
CPAP machine enclosure and create a risk of fire. This applies to most types of 
CPAP machines.
Oxygen supports combustion. Oxygen should not be used while you are
smoking or in the presence of an open flame.
Always ensure airflow is being generated by the device before the oxygen
supply is turned on.
Always turn the oxygen supply off before stopping the airflow from the device.
Note: At a fixed rate of supplemental oxygen flow, the inhaled oxygen concentration will 
vary, depending on where the oxygen is introduced, the pressure settings, patient breathing 
pattern, mask selection, and leak rate.
Do not use the 
S8 E
SCAPE
 if there are obvious external defects or unexplained 
changes in performance.
Do not open the 
S8 E
SCAPE
 case. There are no user serviceable parts inside. 
Repairs and internal servicing should only be performed by an authorized service 
agent.
Explosion hazard—do not use in the vicinity of flammable anesthetics.
C
AUTIONS
At low pressures, the flow through the exhalation ports of your mask may not 
clear all exhaled gas from the tubing. Some rebreathing may occur.
The airflow for breathing produced by this device can be as much as 11
o
F (6
o
C) 
higher than the temperature of the room. Caution should be exercised if the room 
temperature is warmer than 90
o
F (32
o
C).
When AC mains power (100–240V AC) is not available, always use a ResMed 
DC-12 converter. (The DC-12 converter is available as an optional accessory. It is 
not supplied with all models.)
If supplied, the Data Card module is designed for use with specific ResMed 
devices.
Only use the Data Card module (if supplied) as specified in this manual.
Note: The above are general warnings and cautions. Specific warnings, cautions, and 
notes appear with the relevant instructions in the manual.