THEORY OF OPERATION 
Model 360E Instruction Manual 
 
IR 
Source 
IR Beam 
Sample Chamber 
Band-Pass Filter 
Photo-Detector 
 
Figure 10-1:   Measurement Fundamentals 
Upon exiting the sample cell, the beam shines through a band-pass filter that allows only light at a wavelength of 
4.3 µm to pass.  Finally, the beam strikes a solid-state photo-detector that converts the light signal into a 
modulated voltage signal representing the attenuated intensity of the beam.   
10.1.3. Gas Filter Correlation 
Unfortunately, water vapor absorbs light at 4.3 µm too.  To overcome the interfering effects of water vapor the 
Model 360E adds another component to the IR light path called a gas filter correlation (GFC) wheel (see Figure 
10-2). 
Measurement Cell
(Pure N
2
)
 
Reference Cell 
(N
2
 with CO
2
)
 
 
Figure 10-2:   GFC Wheel 
10.1.3.1. The GFC Wheel 
A GFC wheel is a metallic wheel into which two chambers are carved.  The chambers are sealed on both sides 
with material transparent to 4.3 µm IR radiation creating two airtight cavities.  Each cavity is filled with specially 
composed gases.  One cell is filled with pure N
2
 (the measure cell)
.  
The other is filled with a combination of N
2
 
and a high concentration of CO
2
 (the reference cell). 