This appendix provides information on the default values for most Series 3000 parameters. Where
appropriate, valid ranges are also defined.
appropriate, valid ranges are also defined.
These default values represent the platform configuration after a master reset. Depending on how the
platform was ordered, certain values may have been configured at the factory.
platform was ordered, certain values may have been configured at the factory.
E.2
Most frequently used defaults and ranges
The table below contains the default values and ranges for the most frequently used Series 3000
settings.
settings.
Table E-1
Series 3000 default values and ranges
Type
Setting
Default
Range
Comments
Flow
Flow direction
Forward only
Flow damping
0.8 sec
0.0 – 51.2 sec
User-entered value is
corrected to nearest lower
value in list of preset values.
For gas applications, Micro
Motion recommends a value
of 2.56 seconds or higher.
corrected to nearest lower
value in list of preset values.
For gas applications, Micro
Motion recommends a value
of 2.56 seconds or higher.
Flow calibration factor
1.00005.13
Mass flow units
g/s
Mass flow cutoff
0.0000 g/s
Recommended setting:
• Standard use – 0.2% of the
sensor’s rated maximum
flowrate
flowrate
• Empty-full-empty batching –
2.5% of the sensor’s
maximum flowrate
maximum flowrate
Volume flow type
Liquid
Volume flow units
L/s
Volume flow cutoff
0.0000 l/s
0.0 –
x l/s
x is obtained by multiplying
the flow calibration factor by
0.2, using units of l/s
the flow calibration factor by
0.2, using units of l/s
Meter factors
Mass factor
1.00000
Density factor
1.00000
Volume factor
1.00000