Emerson Model 1700 Benutzerhandbuch
Option
Description
Disabled
(default) Operators cannot start and stop totalizers and inventories from the dis-
play.
5.2.2
Enable or disable Totalizer Reset from the display
Display (standard)
OFF-LINE MAINT
> OFF-LINE CONFG > DISPLAY > TOTALS RESET
Chinese-language
display
display
Offline Maintain
> Configuration > Display > Reset Total
ProLink II
ProLink
> Configuration > Display > Display Options > Display Totalizer Reset
ProLink III
Device Tools
> Configuration > Totalizer Control Methods
Field Communicator Configure > Manual Setup > Display > Display Variable Menu Features > Totalizer Reset
Overview
You can configure whether or not the operator is able to reset totalizers from the display.
Restrictions
•
•
This parameter does not apply to inventories. You cannot reset inventories from the display.
•
You cannot use the display to reset all totalizers as a group. You must reset totalizers
individually.
individually.
•
If the petroleum measurement application is installed on your computer, the operator must
enter the off-line password to perform this function, even if the off-line password is not
enabled.
enter the off-line password to perform this function, even if the off-line password is not
enabled.
Procedure
1.
Ensure that the totalizers you want to reset have been configured as display
variables.
variables.
If the totalizer is not configured as a display variable, the operator will not be able to
reset it.
reset it.
2.
Enable or disable resetting the totalizer as desired.
Option
Description
Enabled
Operators can reset a totalizer from the display, if the totalizer is config-
ured as a display variable.
ured as a display variable.
Disabled
(default) Operators cannot reset totalizers from the display.
Configure device options and preferences
Configuration and Use Manual