A&D ML-50 Benutzerhandbuch

Seite von 75
MS-70 / MX-50 / MF-50 / ML-50 
23 
  11.  Open the heater cover. Put in a sample using the 
level indicator. 
 
 
 
 
 
Caution 
 
 
The sample needs to be more than 0.1 g. 
  Spread 
the 
sample 
evenly. 
 
 12.  If the 
 key is pressed during gram display, analysis 
mode, measurement unit and minimum scale value 
are displayed. If the 
 ENTER 
  key is pressed, 
sample mass is output. 
 
 
Star
Star
Star
Start the Measurement
t the Measurement
t the Measurement
t the Measurement    
  13.  Close the heater cover. Press the 
 START 
 key 
after a stable value is displayed. 
 
 14.  If 
the 
 SELECT 
  key is pressed during measurement, 
other units can be displayed temporarily and its 
measurement values can be output. 
 
Caution 
 
 
Do not press the 
 SELECT 
 key while 
sampling data. 
 
 
Finish the Measurement
Finish the Measurement
Finish the Measurement
Finish the Measurement    
  15.  When the change of moisture content per one 
minute (drying rate) reaches the preset termination 
value, the measurement is completed and the 
buzzer beeps. 
 
  16.  Open the heater cover and remove the sample 
using the pan handle. 
 
 
  17.  Press the following key to return to gram display. 
 
 
 ENTER 
  key..... Outputs (Prints) result. 
 
 
 SELECT 
  key ... Returns to the weighing mode. 
 
 
 RESET 
  key ..... Returns to the weighing mode and 
displays zero. 
 
 
  18.  If the same condition is used, proceed to step 8. 
 
  If changing the condition, proceed to step 2. 
 
 
  The sample pans can be washed and reused. 
 
  There is the Reference card on the bottom of the analyzer. 
No good 
Good 
Close
Remove sample 
Drying temperature 
Analysis time 
Drying rate 
Result 
Sample needs more than 0.1 g. 
Significant sample is indicated 
segment. 
Range of proper sample