A&D MS-70 Benutzerhandbuch

Seite von 75
MS-70 / MX-50 / MF-50 / ML-50 
66 
 
 
15.5.
15.5.
15.5.
15.5. Error Message
Error Message
Error Message
Error Message    
 
 CH  no   Internal Error 
 
An internal error indicated by the result of the self-check function. If repair is 
needed, contact the local A&D dealer. 
 
 Cl  pf   Clock Battery Error 
 
Press any key and input the date and time. Refer to "6.2. Setting the Clock 
and Calendar ". 
 
 Cl err   Clock Error 
 
Contact the local A&D dealer to repair the analyzer. 
 
 Clo5e    Heater Cover Error 
 
The heater cover is opened when starting self-check function. If it is closed, 
the self-check function is started. 
 
 error0   Internal Error 
 
Turn the power switch off and then on .   
Check the frequency of the power supply. 
Contact the local A&D dealer to repair the analyzer, if the error is not cleared.   
 
 error3   IC Error 
 error8   Contact the local A&D dealer to repair the analyzer. 
 error9  
 
 Ht err   Temperature Control Error 
 
Contact the local A&D dealer to repair the analyzer, if an error is not cleared 
when turning the power switch off for more than a half hour and rechecking it. 
 
 t-Up     Time Error at Temperature Calibration 
 
There is no key operation for five minutes during temperature calibration. 
 
If pressing any key, the weighing mode is displayed. 
 
   e      Positive Overload, Overweight 
 
The sample has exceeded the weighing capacity. 
 
If the weighing sample pan is empty and this error is displayed, contact the 
local A&D dealer to repair the analyzer. 
 
  -e      Negative OverloadSample Pan Error 
 
The weight value is too light. 
 
Check the pan, pan support and press the 
 RESET 
 key. 
 
Calibrate the weighing sensor. 
 
If an error can be not cleared, contact the local A&D dealer to repair the analyzer. 
 
 Full 
Memory 
 
 
The number of results stored in memory has reached the upper limit. 
 
 
Clear the data to store the new results. Refer to "12. Data Memory Function". 
 
MEM 
 fUl