A&D MX-50 Benutzerhandbuch

Seite von 75
MS-70 / MX-50 / MF-50 / ML-50 
39 
    
 
1.  Display gram unit (of the weighing mode). 
 
Select a Program Number to Edit the Measurement 
Select a Program Number to Edit the Measurement 
Select a Program Number to Edit the Measurement 
Select a Program Number to Edit the Measurement Program
Program
Program
Program    
 2. Press 
the 
 PROGRAM 
  key and press the 
 or 
 key to select a program number. 
 
 3. Press 
the 
 ENTER 
  key to use the number. 
 
 
4.  The analyzer displays 
 end 
  and returns to the 
weighing mode. 
 
Caution 
If the data memory function is active, the data 
memory number (MEM) is displayed in place of the 
program number (PROG).
 
 
 
Select an Analysis Mode
Select an Analysis Mode
Select an Analysis Mode
Select an Analysis Mode    
 5. Press 
the 
 SELECT 
 key to select a symbol, it 
will blink. Select 
 U-t 
 of the timer mode with 
the 
 or 
 key. (Press these keys to select it) 
 
Set the
Set the
Set the
Set the Heating Pattern
 Heating Pattern
 Heating Pattern
 Heating Pattern    
  6.  Select drying temperature with the 
 SELECT 
 key. 
Select step drying 
 
of the heating pattern 
with the 
 PROGRAM 
 key. 
 
   
 
 
Set Drying Temperature 1
Set Drying Temperature 1
Set Drying Temperature 1
Set Drying Temperature 1    
 
7.  Set 120°C with the 
 or 
 key. 
 
 
Set Time 1
Set Time 1
Set Time 1
Set Time 1    
 
8.  Select time 1 with the 
 SELECT 
 key. 
Set 5.0 minutes with the 
 or 
 key. 
 
 
 
Set Drying Temperature 2
Set Drying Temperature 2
Set Drying Temperature 2
Set Drying Temperature 2    
 
9.  Select Drying Temperature 2 with the 
 SELECT 
 key. 
Set 160°C with the 
 or 
 key. 
 
 
Set Time 2
Set Time 2
Set Time 2
Set Time 2    
  10.  Select time 2 with the 
 SELECT 
 key. 
Set 10.0 minutes with the 
 or 
 key. 
 
 
 
PROG 1
 0.000
g
 
PROG "6
 
 
 end
 
PROG 5
 0.000
g
 
Program number 
 
1
 
"1"2"0°C 
"U"-"t
 0.000
g
 
    
 
 
2
 
"1"6"0°C 
 
 
"1"0"5°C 
 
"1"0."0 min 
Drying Temperature 1 
Analysis mode 
Drying Temperature 2 
Step drying 
Time 2 
Time 1 
 
"5."0 min