3M M5560 Benutzerhandbuch

Seite von 8
Light Source:
-
75 watt tungsten/halogen bulb. Average life 4000 cycles at  20 seconds each.
-
Optically filtered light output for peak power in 400-500 nm band.
-
Fused fiber-optic light guide.
-
13 mm and 8 mm diameter optional light guide aperture
Electrical:
-
Electronically regulated bulb voltage
-
106 VA input power 
-
Fuse: See product name plate for correct rating.
-
Model versions for worldwide electrical supply characteristics. See product nameplate for correct
supply voltage.
Electrical versions available: 50/60 Hz
240V; 230V; 220V; 120V; 110V; 100V
Note: Be sure your unit is correctly rated for your specific electrical supply
-
Unit is double insulated and uses plug configurations according to individual primary 
country specifications.
Protective features:
-
Thermal overload protector in handpiece.
-
Optical filter blocks light outside peak curing band.
-
Protection against electrical shock (Type BF)
-
Double insulated power supply 
-
Low voltage handpiece circuit.
-
Protection against ingress of water: ordinary 
Transport and Storage Conditions:
-
Ambient temperature range -40° C to +70° C
-
Relative humidity range of 10% to 80%, excluding condensation
-
Atmospheric pressure of 500 hPa to 1060 hPa
Dimensions:
-
Power module: 2.7 in. x 6.8 in. x 5.5 in. (69 mm x 173 mm x 141 mm) (2.7kg) 
-
Handpiece: 6.6 in. x 2.1 in. x 5.5 in. (167 mm x 53 mm x 139 mm) (0.3kg) 
Safety Regulations:
M5560 fulfills the following standards:
-
UL Standard 2601-1 /CSA C22.2 No. 601.1/EN60601-1
-
Other country specific deviations to IEC601-1 through CB certification.
Electro Magnetical Compatibility
M5560 complies with IEC 601-1-2 Draft 2
No person is authorized to provide any information which deviates from the information provided in this
instruction sheet.
Warranty
3M ESPE warrants this product will be free from defects in material and manufacture for a period of two
years from date of purchase. 3M ESPE MAKES NO OTHER WARRANTIES INCLUDING ANY IMPLIED
WARRANTY OF MERCHANTABILITY OR FITNESS FOR A PARTICULAR PURPOSE. User is
responsible for determining the suitability of the product for user's application. If this product is defective
within the warranty period, your exclusive remedy and 3M ESPE's sole obligation shall be repair or
replacement of the unit if the unit was new when it was purchased and it is promptly returned to the
following address:
3M Health Care Service Center 
Suite 200, Bldg. 502
3350 Granada Ave. North
Oakdale, MN 55128
800 292 -6298
This warranty does not cover lamps and fiber optic light guides.
Limitation of Liability
Except where prohibited by law, 3M ESPE will not be liable for any loss or damage arising from this
product, whether direct, indirect, special, incidental or consequential, regardless of the theory asserted,
including warranty, contract, negligence or strict liability.
8