9-128 Command Reference
∗RMC
PM6680B/81/85
8 ‘<Macro name>’
Delete one Macro
This command removes an individual MACRo.
Parameters:
‘<Macro name>’ is the name of the macro you want to delete.
+
<Macro name> is String data that must be surrounded by quotation marks.
See also:
*PMC, if you want to delete all macros.
*RST
PM6680B/81/85
Reset
The Reset command resets the counter. It is the third level of reset in a 3-level re-
set strategy, and it primarily affects the counter functions, not the IEEE 488 bus.
The counter settings will be set to the default settings listed on page -. All previous
The counter settings will be set to the default settings listed on page -. All previous
commands are discarded, macros are disabled, and the counter is prepared to
start new operations.
Example:
*RST
See also:
Default settings on page -.
Complies to standards:
IEEE 488.2 1987.