Agilent Technologies A.18.00 Benutzerhandbuch

Seite von 581
288
S:\agilent\8920\8920b\PRGGUIDE\BOOK\CHAPTERS\advoper.fb
 
Chapter 5, Advanced Operations
Status Reporting
Clearing the Event Register   
 
The EVENT register is cleared whenever it is queried or whenever the Common 
Command *CLS is sent to the Test Set. 
Reading the Enable Register   
 
Syntax
STATus:HARDware1:ENABle?
Example
OUTPUT 714;"STAT:HARD1:ENAB?"
ENTER 714;Register_value
Writing the Enable Register   
 
Syntax
STATus:HARDware1:ENABle <integer>
Example
OUTPUT 714;"STAT:HARD1:ENAB 256"
Clearing the Enable Register   
 
The ENABLE register is cleared by writing to it with an integer value of zero.