Changing the Patient Category
Setting up your Monitor   3-51
S
e
tt
ing up y
our
 
M
onit
o
r  
Heart Rate 
(HR) / Pulse
The expected range can be estimated by performing a manual pulse 
count.
RESP
In general, use the Adult or Pediatric category if rates below 50 rpm are 
expected; use Neonatal if expected rate is above 100.
Pressure
The algorithm responsible for respiration artifact suppression is used in 
all three patient categories for the 10 and 30 scales. In the Pediatric and 
Adult categories, the algorithm is also used for the 60 scale.
SpO
2
Default for averaging time is 10 seconds for Adult, Pedi and Neo. 
Note—
Default can be changed in Configuration Mode.
Adult (bpm)
Pedi/Neo (bpm)
Cardiotach Range
15 to 300
15 to 350 
HR Alarm Limit Range 
15 to 250 
15 to 300 
Pleth/Pressure Pulse Alarm Limit 
Range
30 to 250 
30 to 300 
Adult/Pedi (rpm)
Neo (rpm)
Respirotach Range 
0 to 120 
0 to 180 
High Alarm Limit Range  10 to 100
30 to 150
Low Alarm Limit Range
0 to 95
0 to 145