Alarm Display
Alarm Functions   5-3
Ala
rm Func
tions
  
Alarm Lamps
Red and yellow alarm lamps on the control panel signify the severity of 
the cause of the alarm. One of these lamps is always lit in the event of an 
alarm sound. In the case of more than one alarm, the highest severity 
alarm lamp is always lit. 
What the 
 Symbol Means
When the alarm capability is suspended by using the 
 
softkey, or the 
 hardkey, the 
 sign on the control panel is 
illuminated. If, however, you have turned the individual parameter alarms 
off one by one, the 
 sign on the control panel is not illuminated, but a 
 
appears next to the individual parameter numerics on the screen.
Suspending 
Alarms
There are four alarms-off modes available. You configure which one you 
want at installation of your system. The four modes are:
1. ALARMS SUSPND 1 MIN 
2. ALARMS SUSPND 2 MIN 
3. ALARMS SUSPND 3 MIN 
4. ALARMS SUSPENDED
ALARMS SUSPENDED 1, 2, 3 MIN
In these modes the alarm capability can be switched off, but is 
automatically switched on after 1, 2 or 3 minutes (unless you have 
switched it on manually beforehand with the 
 softkey 
in the Alarms Selection Window, or the 
 hardkey).
ALARMS SUSPENDED
In this mode the alarm capability is switched off permanently and must be 
switched on manually when required (using the 
 
softkey in the Alarms Selection Window, or the 
 hardkey).
Note—
For safety reasons, we recommend that you always configure your 
monitor to ALARMS SUSPND 1, 2 or 3 MIN for Intensive Care as well as 
Operating Room use. If you want to avoid nuisance alarms caused by 
artifacts in the Operating Room, configure your alarm limits carefully or 
switch off individual paramter alarms, applying special vigilance to these 
parameters. All monitors within a unit should have the same alarm 
configuration.
Suspend Alarms
Suspend
 
SwitchOn Alarms
Suspend
 
SwitchOn Alarms
Suspend