Installation Information
11-8     Monitor Installation and Patient Safety  
M
onitor
 Ins
ta
lla
ti
on 
a
nd Pa
ti
e
n
t Sa
fe
ty
  
Combining 
Equipment
If it is not evident from the instrument specifications whether a 
particular instrument combination is hazardous or not, for example due 
to summation of leakage currents, the user should consult the 
manufacturers concerned or else an expert in the field, to ensure that 
the necessary safety of all instruments concerned will not be impaired 
by the proposed combination.
Warning
Warning
•Any combination of the Philips M1165/66/67/75/76/77A CMS or 
the Philips M1205A V24 and V26 Patient Monitor with non-
medical equipment must comply with IEC 601-1-1. The monitor 
must be properly grounded, or the non-medical device (e.g. a 
printer or an ITE display) properly isolated. Isolation may be 
achieved by powering the non-medical device through a safety 
isolation transformer. This must be done in order to prevent the 
possibility of an electrical shock hazard (to the user and or 
patient) from excessive ground leakage currents from the non-
medical device such as a printer.
• V24 and V26 only: In normal AC line operation grounding the 
monitor is accomplished by ensuring that the AC line cord is 
connected from the monitor to a properly grounded AC 
receptacle. When the V24CT or V26CT is operated in battery 
mode, the possibility of an electrical shock hazard is especially 
prevalent. Alternatively and additionally, proper grounding may 
be achieved by attaching a safety grounded wire to the threaded 
safety ground (protective earth) stud 
 found on the rear of 
the monitor.