Maintenance Checks
Before commencing monitoring on a patient, carry out the following 
checks on the system:
Check for any mechanical damage
Check all the external leads, plug-ins and accessories
Check all the functions of the instrument which will be needed to 
monitor the patient, and ensure that the instrument is in good 
working order.
Do not use the system for any monitoring procedure on a patient
 
if you identify features which demonstrate impaired functioning of the 
instrument. Contact the hospital biomedical engineer or the Philips 
Service Engineer.
Maintenance
Frequency
Source of Information
Full performance checks 
including safety checks.
At least every two  years,  after 
any repairs, or as needed.
See “Testing and Maintaining the 
CMS” in the “CMS Service Guide” 
or “Testing and Maintaining the 
V24 and V26“in the “V24 and V26 
Service Guide”.
Synchronization of the monitor 
and defibrillator.
If defibrillator is used or 
configured for intended use: at 
least once a year, after any 
repairs, or as needed.
See the Philips M1722/3/4A 
Codemaster XL/XL+/XE 
Defibrillator Service Guide.
Preventive maintenance
Perform the NBP and Sidestream 
CO 2 Module Performance test 
blocks every year.
Perform the Preventive 
Maintenance Procedure for the 
M1180A #A01/A02 Wall Mount 
Assembly every year.
See “Testing and Maintaining the 
CMS” in the “CMS Service Guide” 
or “Testing and Maintaining the 
V24 and V26“in the “V24 and V26 
Service Guide”.