Philips V26CT Benutzerhandbuch

Seite von 392
Attaching the Patient
2-10     Getting Started  
Ge
tti
ng St
a
rte
d  
Power 
Failure
If a power failure occurs, any alarm conditions that were stored in the 
monitor will NOT be retained when power is restored. Settings such as 
alarm limits stored in “T” modules ARE retained - see “Parameter 
Settings Transfer” on page 3-61
.
Caution
You are advised to check all the monitor settings after any power failure.
The monitor reacts according to the length of the power failure:
Patient 
Information 
Center
If the monitor is connected to a Patient Information Center (78504 and 
78508) and you select the same lead for ECG channels 1 and 2, the wave 
is not displayed in channel 2 of the Patient Information Center.
Monitor 
Standby
If you want to stop patient monitoring, but keep all settings and Patient 
Data Management information, you can switch the monitor into 
“Monitor Standby”:
1. Press  
or 
2. Press . 
All waves and numerics disappear from the display, and all monitoring is 
suspended. The word “Standby” appears in big letters on the screen.
Less than 60 seconds
Patient Data Management data and all 
settings retained.
Less than 3 hours
Patient Data Management data retained. All 
settings revert to user defaults.
More than 3 hours
Patient Data Management data not retained. 
All settings revert to user defaults. Time 
setting may be lost if not networked. 
Alarms / Volume
Alarms 
Monitor Standby
Pts
Refernce