oxyCRG Display
Setting up your Monitor   3-25
S
e
tt
ing up y
our
 
M
onit
o
r  
oxyCRG Display
The oxyCRG (oxygen CardioRespiroGram) display provides a 
simultaneous presentation of the following three waveforms overlayed on 
the Main Screen:
Beat-to-beat heart-rate trend
An oxygen parameter trend
Compressed respiration wave
Approximately the last six minutes of information is displayed for each 
parameter.
Main Screen Display with oxyCRG
The oxyCRG display can help you to make early detections of respiratory 
regulation problems in neonates, by comparing trend patterns, or finding 
interrelationships among the three parameters.