NGA-2000 System Calibration
90003482(2) [NGA-e (MLT-Software 3.2.X)] 07/98
NGA 2000
Supplement - 9
3.2 Programming Calibration Sequences
Beside the standard programs „zero calibration“ and „zero/span calibration“ it is possible to
run the system calibration in an user definable order of up to 40 steps.
run the system calibration in an user definable order of up to 40 steps.
The menu to set-up this feature looks as follows.
SCSTEP
SCCALTYPE1
SCCALTYPE1
SCSTEPMOD
For programming the sequence you must
1. select „Step #“
2. select „Calibration procedure type“
3. select analyzer module/ MLT channel
2. select „Calibration procedure type“
3. select analyzer module/ MLT channel
Repeat this order for all program steps.
The programmable „Calibration procedure types“ are:
1. NoOp
no operation (for deleting a step in an existing program)
2. Zero-Cal
do a zero calibration
3. Span-Cal
do a span calibration for all available ranges
4. Zero/Span-Cal
do a zero cal then a span cal for all available ranges
5. Span1-Cal
do a span calibration only for range #1
6. Span2-Cal
do a span calibration only for range #2
7. Span3-Cal
do a span calibration only for range #3
8. Span4-Cal
do a span calibration only for range #4
9. END-OF-PGRM
end of sequence
System configuration and diagnostics...
↓
System calibration...
↓
Calibration sequence programming...
↓
-- Calibration Sequence Programming --
Step #: 1
Calibration procedure type: ZERO-CAL
Choose specific analyzer module...
Select all analyzer modules!
Program steps 1-10...
Program steps 11-20...
Program steps 21-30...
Program steps 31-40...
_______________________________________________
Analyzer-module:
Calibration procedure type: ZERO-CAL
Choose specific analyzer module...
Select all analyzer modules!
Program steps 1-10...
Program steps 11-20...
Program steps 21-30...
Program steps 31-40...
_______________________________________________
Analyzer-module:
MLT/CH3