Epson 9840 Benutzerhandbuch

Seite von 59
I N T E R F AC E   M O D E L   9 8 4 0   V E R   6 . 0 . 2  
 
2. Full = 4.0000 Lb 
Zero = 4.0000 Lb 
 
(Input) Mv/V 
(Output)Volt 
(Output) MA 
 
 
 
4 0v 
12mA 
3 -2.5v 
10mA 
2 -5.0v 
8mA 
1 -7.5v 
6mA 
0 -10.0v 
4mA 
-1 -10.0v 
4mA 
-2 -10.0v 
4mA 
-3 -10.0v 
4mA 
-4 -10.0v 
4mA 
 
3. Full = 2.0000 Lb 
Zero = 0.0000 Lb 
 
(Input) Mv/V 
(Output) Volt 
(Output) MA 
 
 
 
4 10.0v 
20mA 
3 10.0v 
20mA 
2 10.0v 
20mA 
1 5.00v 
16mA 
0 0v 
12mA 
-1 -5.0v 
8mA 
-2 -10.0v 
4mA 
-3 -10.0v 
4mA 
-4 -10.0v 
4mA 
 
4. Full = 2.0000 Lb 
Zero = 2.0000 Lb 
 
(Input) Mv/V 
(Output) Volt 
(Output) MA 
 
 
 
4 10.0v 
20mA 
3 5.0v 
16mA 
2 0v 
12mA 
1 -5.00v 
8mA 
0 -10.00v 
4mA 
-1 -10.0v 
4mA 
-2 -10.0v 
4mA 
-3 -10.0v 
4mA 
-4 -10.0v 
4mA 
 
>Sensor Select 
This menu is used to manually select load or torque cell calibration data from a list stored 
in the Model 9840.  When you select this menu on a 2 channel unit, you are asked to 
select Channel A or B.  The cell serial number of the selected channel is displayed.  Use 
either the left/right or plus/minus buttons to page through the list.  The escape button will 
M O D E L   9 8 4 0                                                                                                                                                          P G   1 7  
P U B .
 
2 8 5 6 - 1 6