I N T E R F AC E   M O D E L   9 8 4 0   V E R   6 . 0 . 2  
set the torque.  When the Model 9840 is reading the cell it will display the “Reading...” 
message for about 10 seconds.  During this time it is important that the masses or 
torque are not disturbed since all the readings taken are averaged together to obtain 
the calibration data.  The second mass/torque is then entered and read.  If 5 Point 
calibration was selected, masses/torque three through five will be entered and read.  
One of these readings should be at zero.  A "No Masses Ready?” or "No Torque 
Ready" reading is also required.  During this reading, a shunt check value is recorded.   
    >>Cal by Shunt Val 
This item is used to calibrate a cell using the manufacturer’s shunt calibration factor 
expressed in pounds or pound-inches.  You will be asked for the channel, the cell 
serial number, the date, the calibration unit, the cell’s maximum rated load or torque, 
the desired excitation voltage (5.0 or 10.0 volts), and the shunt calibration constant.  
The Model 9840 will then automatically perform the shunt calibration of the cell using 
its internal precision shunt resistor.  Note that there is a back panel switch that allows 
you to select either a 30K or 60K resistor for shunt measurements (see Appendix B for 
a circuit diagram).  Be sure to select the appropriate value for your cell.  During this 
calibration the “Reading...” message will be displayed.  Do not disturb the cell while it 
is being read.  Like the “Cal by mV/volt” option, the cell should be free of load or 
torque during this calibration. 
 
    >>Counts per Inch 
This entry allows you to set the number of counts per inch that the quadrature encoder 
input will use. 
>Limits 
This menu is used to set up the 4 contact closure switches.  Instead of a submenu there 
are a series of selections.  Pressing Enter will accept the displayed entry.  Pressing 
Escape will return to the setup mode main menu without changing the limit setup. Note 
that changing a cell type on a channel will automatically erase limits that were set for that 
channel.  These limits must then be reset by the user. 
 
1.  Select Limit X -- The first selection chooses which of the limits you want to set up (1-
4).  Use the left/right or plus/minus buttons to change the number.  Press enter to 
accept the displayed number.  Press escape to return to the setup mode main menu. 
2.  Normally Open / Closed -- This selection determines if the switch contacts are open 
or closed when the limit is de-activated. 
3.  Enabled / Disabled -- Choosing enable allows the limit to operate as specified by the 
following selections.  Choosing disable places the limit in its “normal” state and 
completes the limit setup. 
4.  Source:  -- This selection allows you to scroll through a list (use either the left/right or 
plus/minus buttons) to select the source of the data that will drive the limit.  The 
standard list will include load or torque, peak, valley, position, and velocity.  Press 
enter to select the displayed item. 
5.  Units: -- This entry allows you to scroll through the list of available units for the data 
source you have chosen in the previous step. 
6.  Set  -- This is the numerical value of the source at which the limit will be activated (set 
point).  It is specified in the units you chose in the previous step. 