I N T E R F AC E   M O D E L   9 8 4 0   V E R   6 . 0 . 2  
RS232 SETUP MODE COMMANDS 
The commands in this section allow you to change the setup of the Model 9840.  You will 
notice that these commands form a menu which approximately duplicates the Setup Mode 
menu used during setup of the Model 9840 from the front panel.  There are additional 
commands that are for RS232 use only. 
 
If the optional Keylock is installed and locked you will not be to alter the setup until the unit 
has been unlocked.  The acknowledgement will indicate this fact: 
@123  Setup is Locked - use Keylock to unlock 
User Data Entry Command (U) 
This command is used to view or set the options in the user data entry menu, namely the 
base area (used to calculate load in PSI, or MPa) and base length (used to calculate 
position in %).  This command has three subcommands: User data View (UV), User data 
Area (UA), and User data Length (UL). 
    User Data View Command (UV) 
The User data View (UV) command has the format UV. 
Example: 
@123UV 
Acknowledge: 
@123 Base Area Ch A is 1.00000 sq-in 
     Base Area Ch B is 0.50000 sq-in 
     Base Length is 3.0000 in 
    User Data Area Command (UA) 
The User data Area (UA) command is used to set the base area.  The format is 
UA(channel)(number)#. 
Example: 
@123UAA1.0025# 
Acknowledge: 
@123 Base Area Ch A is 1.0025 sq-in 
    User Data Length Command (UL) 
The User data Length (UL) command is used to set the base length.  The format is 
UL(number)# 
Example: 
@123UL2.5# 
Acknowledge: 
@123 Base Length is 2.5000 in 
M O D E L   9 8 4 0                                                                                                                                                          P G   3 2  
P U B .
 
2 8 5 6 - 1 6