I N T E R F AC E   M O D E L   9 8 4 0   V E R   6 . 0 . 2  
Calibration Escape Command (CE) 
The Calibrate Escape (CE) command is used to cancel a calibration that was started 
with the Calibrate Begin (CB) command.  This format is CE. 
Example: 
@123CE 
Acknowledge: 
@123 Calibrate Command - Canceled, Calibration NOT Changed 
Calibrate by Millivolt-Per-Volt Command for One Point (CV) 
The Calibrate by milli-volt per Volt (CV) command is used to calibrate a load cell using 
its millivolt per volt calibration constant.  You must have already started a calibration 
using the CB command for this command to function.  The format is CV(millivolt per 
volt constant)#  Note that a shunt check calibration will also be run at this time and so it 
is important that the load cell is not under load when this command is entered. Note 
that this command works for torque cell calibration as well. 
 
Example: 
@123CV4.5002# 
Acknowledge: 
@123 Calibrate Command - Reading for Shunt Check... 
          [10 seconds delay] 
@123 Calibrate Command Completed 
     Ch A = S/N 123456, 1000.0 Lb      , 4.50020 mV/v,  
              10.00 V   , Cal on Oct27-99, 553.26 Lb  Shunt 
     unused S/N  63220, 100.00 Lb      , 2.99984 mV/v,  
            10.00 V   , Cal on Oct27-99, 49.381 Lb  Shunt 
     Ch B = S/N  89991, 1000.0 Lb      , 4.50015 mV/v,  
             5.00 V   , Cal on Oct27-99, 486.45 Lb  Shunt 
Calibrate by Millivolt-Per-Volt Command for Six Points (CMV6) 
The Calibrate by milli-volt per Volt for six points (CMV6) command is used to calibrate 
a load cell using 6 points of mass and milli-volt per volt each.  You must have already 
started a calibration using the CB command for this command to function.  The format 
is CMV6. Note that a shunt check calibration will also be run at the end of the 
calibration and so it is important that the load cell is set at 0 (or no masses) when the 
final command is entered. 
 
 
Example: 
@123CMV6 
Acknowledge: 
@123 Calibrate by mV/Volt - 6 Point 
 
Ready for Mass CMVM1 command 
Calibrate by mV/V Mass and Volt Commands (CMVM and CMVV) 
The Calibration by mV/V Mass (CMVM) and Volt (CMVV) commands can have up to 
seven parts each: CMVM1 through CMVM6 and CMVM0; and CMVV1 through CMVV6 
and CMVV0.  You must have already started a calibration using the CB and CMV6 
commands for any of these commands to function.  To input the first mass issue the 