Epson 9840 Benutzerhandbuch

Seite von 59
I N T E R F AC E   M O D E L   9 8 4 0   V E R   6 . 0 . 2  
    Display Count By Command (DC) 
The Display Count by (DC) command sets the 
smallest number that the display will count by 
for the Load or Torq, Peak, and Valley 
measurements.  The format is 
DC(channel)(code) where the code is selected 
from the table at the right.  
Display Count by Command 
Code Count 
By 
0 1 
1 2 
2 5 
3 10 
4 20 
Example: 
@123DCA3 
Acknowledge: 
@123 Channel A counts by 10 
    
 Display Second Line Command (D2) 
The Display Second Line (D2) command is used to set the second line to be blank, or 
to show the limit status.  The format is D2(B, L, D, or T), where B makes the line blank, 
L makes the line show limit status, D makes the line show another display, and T 
makes the line show text. 
Example: 
@123D2B 
Acknowledge: 
@123 Second Line set to Blank 
    Display Second Line Text (DT) 
The Display Second Line Text (DT) command is used to select the text that would be 
displayed on the second line if text was set to go on the second line. 
Example: 
@123DTModel 9840 
Acknowledge: 
@123 Text Message – Model 9840   
 
   Display Filter Window Commands (DW) 
The Display Filter Window (DW) command set enables and disables the filter window 
and allows the user to enter the unit and value of the filter window to use.  There are 
two commands used to enable and set the filter windows. The format to enable or 
disable the Filter Window is DW1(channel)(0 or 1) where channel is A or B, 1 enables 
the window and 0 disables the window. The format to set the window unit and value is 
DW2(channel)(window unit)(window value)# where window unit is the desired unit for 
the filtering window, and window value is the desired value of the window. The units 
available will depend on the sensor that is currently active for the channel.  It is a good 
idea to issue the @123? command and find out what the active sensor is on the 
channel with the @123SA or @123SB commands to make sure that the appropriate 
unit will be entered.  The window must be enabled before a value can be entered.  
M O D E L   9 8 4 0                                                                                                                                                          P G   4 9  
P U B .
 
2 8 5 6 - 1 6