Section 3: Installation
February 3, 2000
70055MP Revision D
Alarm Configuration Jumpers
Jumpers JP18 and JP19 set the annunciator relay outputs to normal or
inverted. Jumper JP13 configures the
Acknowledge Alarm
button. Refer to
Table 3-3 for jumper positions. Refer to Figure 3-3 for jumper locations.
NOTE: JP13 must always be installed.
Table 3-3: Alarm Configuration
Jumper
Position
Description
JP18
INV
NORM
NORM
Annunciator 1 Output Inverted.
Annunciator 1 Output Normal.
Annunciator 1 Output Normal.
JP19
INV
NORM
NORM
Annunciator 2 Output Inverted.
Annunciator 2 Output Normal.
Annunciator 2 Output Normal.
JP13
Installed
Removed
Removed
Configures the Alarm Acknowledge button for use.
Not applicable.
Not applicable.
Ground Jumpers
Jumpers JP10 and JP14, when installed, are set to earth ground. These
jumpers must always be installed for accurate and safe operation.
WARNING: Jumpers JP10 and JP14 must always be installed for accurate
and safe monitor operation.