SITRANS LU01 – INSTRUCTION MANUAL
7ML19985BE02
Pa
ra
m
e
te
rs
To manually record a profile
Press TRANSDUCER 
 to fire the transducer and record the echo profile into the 
internal scope buffer for display.
For differential or average operation (P001 = 4 or 5), access Scope Displays (P810) 
parameter to select the transducer number.
To save a manual record
Press ENTER 
 to copy the echo profile record in the scope buffer and save it in the 
selected address in the record library. The parameter value field displays the new record 
information.
To display a record
Press 
 key to enter display auxiliary mode.
Then press TRANSDUCER 
 to copy the current echo profile into the scope buffer for 
display on an oscilloscope or Dolphin Plus
To delete a record
Press CLEAR 
 and then ENTER 
 to delete the echo profile record in the selected 
address. The value returns to - - - -.
P331 Auto Record Enable
Use to enable/disable the Auto Profile Record function. 
P333 Auto Record Interval
Enter the time to elapse after an Auto Profile Record is saved before another Auto Profile 
Record can be saved (subject to all other restrictions).
Auto Record ON and OFF Setpoints (P334 to P337)
Use Auto Record ON Setpoint (P334) and Auto Record OFF Setpoint (P335) to define the 
boundaries within which the level must be, for the resultant Echo Profile to be considered 
for an Auto Profile Record.
If ---- is displayed for either P334 or P335, Auto Profile Records are saved regardless of 
current level (subject to all other restrictions).
Enter the level value in Units (P005) or percent of Span (P007) as referenced to 
Empty (P006). 
Values
Range: 0 to 1
0
*
Off
1
On
Values
Range: 0.0 to 9999 (minutes)
Preset: 120