Intr
oduct
io
n
Introduction 
SITRANS LU01 
The SITRANS LU01 is an ultrasonic long-range level monitoring system for liquids and 
solids. SITRANS LU01 level monitor uses one Siemens Milltronics ultrasonic transducers 
(ordered separately) to accurately monitor material levels without material contact. 
The SITRANS LU01 transmits electronic pulses to each connected ultrasonic transducer. 
The transducer converts the electronic pulses to ultrasonic pulses which are emitted 
from the transducer face in a narrow beam. The SITRANS LU01 measures the time from 
the pulse emission, to reception of the reflection (echo) from the material. Using the time 
measured, the SITRANS LU01 calculates the distance from the transducer face to the 
material. 
The distance calculation depends upon the sound velocity within the vessel. When an 
Echomax transducer is used, variable air temperatures are automatically compensated. 
For superior air temperature compensation, a Siemens Milltronics TS-3 temperature 
sensor may be used for each vessel. A simple calibration feature compensates for 
homogeneous atmospheres other than air.
The SITRANS LU01 uses Siemens Milltronics patented Sonic Intelligence®echo 
processing. Sonic Intelligence provides high measurement reliability, regardless of 
changing conditions within the vessel monitored. By using ultrasonic echo ranging 
principles with Sonic Intelligence and velocity compensation, the SITRANS LU01 provides 
outstanding measurement accuracy, usually within 0.25% of range. 
Note: 
The SITRANS LU01 is to be used only in the manner outlined in this instruction 
manual. 
This product is intended for use in industrial areas. Operation of this equipment 
in a residential area may cause interference to several frequency based 