National Instruments NI 4070/4072 6 1/2-Digit FlexDMM Benutzerhandbuch

Seite von 68
NI 4070/4072 Calibration Procedure
60
ni.com
500 mV
100 kHz
500 mV
AC/DC
0.4974 V
0.5026 V
500 mV
300 kHz
500 mV
AC/DC
0.48475 V
0.51525 V
5 V
30 Hz
5 V
DC
4.9945 V
5.0055 V
5 V
50 Hz
5 V
AC/DC
4.9965 V
5.0035 V
5 V
1 kHz
5 V
AC/DC
4.9965 V
5.0035 V
5 V
1 kHz
50 V
AC/DC
4.9875 V
5.0125 V
5 V
1 kHz
300 V
AC/DC
4.9375 V
5.0625 V
5 V
20 kHz
5 V
AC/DC
4.9965 V
5.0035 V
5 V
50 kHz
5 V
AC/DC
4.9945 V
5.0055 V
5 V
100 kHz
5 V
AC/DC
4.974 V
5.026 V
5 V
300 kHz
5 V
AC/DC
4.8475 V
5.1525 V
50 V
30 Hz
50 V
DC
49.945 V
50.055 V
50 V
50 Hz
50 V
AC/DC
49.965 V
50.035 V
50 V
1 kHz
50 V
AC/DC
49.965 V
50.035 V
50 V
20 kHz
50 V
AC/DC
49.965 V
50.035 V
50 V
50 kHz
50 V
AC/DC
49.945 V
50.055 V
50 V
100 kHz
50 V
AC/DC
49.74 V
50.26 V
50 V
300 kHz
50 V
AC/DC
48.475 V
51.525 V
219 V
30 Hz
300 V
DC
218.751 V
219.249 V
219 V
50 Hz
300 V
AC/DC
218.8305 V
219.1695 V
219 V
1 kHz
300 V
AC/DC
218.8305 V
219.1695 V
219 V
20 kHz
300 V
AC/DC
218.8305 V
219.1695 V
219 V
50 kHz
300 V
AC/DC
218.7429 V
219.2571 V
219 V
100 kHz
300 V
AC/DC
217.845 V
220.155 V
70 V
300 kHz
300 V
AC/DC
67.75 V
72.25 V
Table 16.  NI 4070/4072 AC Voltage Verification Limits (Continued)
Calibrator Output
Range
Coupling
2-Year Limits
 Amplitude
 Frequency
Lower
Upper