Campbell Hausfeld SDM-CVO4 Benutzerhandbuch

Seite von 30
SDM-CVO4 4-Channel Current/Voltage Output Module 
4:  Volt (SE) (P1) 
 1:  1 
Reps 
  2:  32 
50 mV, 50 Hz Reject, Slow Range 
 3:  3 
SE 
Channel 
  4:  4 
Loc [ Radiation ] 
 5:  100 
Mult 
 6:  0.0 
Offset 
 
; Now output data in the normal way to final storage 
 
5:  If time is (P92) 
  1:  0 
Minutes (Seconds --) into a 
  2:  1 
Interval (same units as above) 
  3:  10 
Set Output Flag High (Flag 0) 
 
6:  Real Time (P77) 
  1:  110 
Day,Hour/Minute (midnight = 0000) 
 
7:  Wind Vector (P69) 
 1:  1 
Reps 
  2:  0 
Samples per Sub-Interval 
  3:  0 
S, é1, & å(é1) Polar 
  4:  1 
Wind Speed/East Loc [ WindSpeed ] 
  5:  2 
Wind Direction/North Loc [ WindDir   ] 
 
8:  Average (P71) 
 1:  2 
Reps 
  2:  3 
Loc [ AirT      ] 
 
; Now the code to scale the values and update the SDM-CVO4 
; As we have four channels to output, we will first copy the  
; current readings with P54, block move, in a block of four 
; input locations to hold the scaled outputs 
 
9:  Block Move (P54) 
  1:  4 
No. of Values 
  2:  1 
First Source Loc [ WindSpeed ] 
 3:  1 
Source 
Step 
  4:  5 
First Destination Loc [ ScldOut_1 ] 
 5:  1 
Destination 
Step 
 
; Then apply the scaling with one Instruction P53 
; The readings are scaled -5000 to +5000, i.e. to 
; cover the full scale range which would equate to 
; 0 - 10,000 mV in voltage mode or 0-20 mA in  
; current mode. 
 
; Windspeed to cover the range 0-100 m/s 
; Wind direction to cover 0-360 degrees 
; Temperature -25 to +50 degrees C 
; Radiation 0 - 1000  m^2/s 
 
10:  Scaling Array (A*Loc+B) (P53) 
  1:  5 
Start Loc [ ScldOut_1 ] 
  2:  100 
A1 ; WS multiplier 
  3:  -5000 
B1 ; WS Offset 
 4:  27.7778  A2 
 5:  -5000 
B2 
 6:  133.333  A3 
16