A&D AD-4212A/B Benutzerhandbuch

Seite von 102
 
30
 
Class Item 
and 
Parameter 
Description 
 0 
3-level comparator contact 
output 
Cpout 
Comparator 
output 
Cp-t 
Comparator output 
setting 
 1 
5-level comparator contact 
output 
Only the AD-4212B series 
balance displays this class and 
item.
 
 0 
No comparison 
 1 
Comparison, excluding “near zero” when stable or overloaded 
 2 
Comparison, including “near zero” when stable or overloaded 
 3 
Continuous comparison, excluding “near zero” 
Cp 
Comparator mode
 
 4 
Continuous comparison, including “near zero” 
 0 
Not added 
Cp-r 
*2
 
Comparison results 
 1 
Added 
Use in A&D standard format. 
 0 
OFF 
bep
 
LL buzzer 
1
ON
Displayed only when 5-level 
comparator output (
*3
) is selected. 
0
OFF
bep_
 
LO buzzer 
1
ON
 
0
OFF
bep-
 
OK buzzer
1
ON
 
 0 
OFF 
bep
 
HI buzzer 
1
ON
 
 0 
OFF 
Cp fnc 
Comparator
 
bep
 
HH buzzer 
 1 
ON 
Displayed only when 5-level 
comparator output (
*3
) is selected. 
Cp HH 
Secondary upper limit
 
Cp Hi 
Upper limit
 
Cp lo 
Lower limit
 
Cp ll 
Secondary lower limit
 
Refer to "10-10 Comparator 
Function”. 
Displays Cp HH,, Cp ll only when 
5-level comparator output (
*3
) is 
selected. 
 0 
Key mode (when stable) 
Accepts the PRINT key or external 
contact input to output (or store) 
data only when the display is 
stable. 
 1 
Auto print mode A 
(Reference = zero) 
 2 
Auto print mode B 
(Reference = last stable value) 
Outputs (or stores) data when the 
display is stable and conditions of   
ap-p, ap-b and the reference value 
are met. 
 3 
Stream mode /   
Interval memory mode 
With data 0, outputs data 
continuously; with data 2, uses 
interval memory. 
 4 
Key mode B (instantly) 
Accepts the PRINT key or external 
contact input to output (or store) 
data, regardless of the display 
condition. 
 
 
 
 
dout 
Data output
 
 
Not displayed 
when BCD 
output (OP-01) 
is installed.
 
 
 
 
 
 
 
prt
 
Data output mode
 
 5 
Key mode C (when stable) 
Accepts the PRINT key or external 
contact input to output (or store) 
data, instantly when stable or after 
the display is stabilized. 
 
 
Factory setting 
 
*2  This item is not displayed for the AD-4212B series balance or when BCD output (OP-01) is 
installed. 
*3  AD-4212A series balance: when OP-04 is installed 
 
AD-4212B series balance: when “Cp-t” is set to “1”