18. COMMANDS 
18-1 Command List 
Commands to query weighing data 
C Cancels 
the 
S or SIR command. 
Q 
Requests the weighing data immediately. 
S 
Requests the weighing data when stabilized. 
SI 
Requests the weighing data immediately. 
SIR 
Requests the weighing data continuously. 
 
Commands to control the balance 
CAL 
Same as the CAL key.   
OFF 
Turns the display off. 
ON 
Turns the display on. 
P 
Same as the ON:OFF key 
PRT 
Same as the PRINT key 
R 
Same as the RE-ZERO key 
SMP 
Same as the SAMPLE key. 
U 
Same as the MODE key (Confirm the upper/lower limit values) 
HH:
∗∗∗∗∗∗∗∗      g 
H I :
∗∗∗∗∗∗∗∗      g 
LO:
∗∗∗∗∗∗∗∗      g 
LL :
∗∗∗∗∗∗∗∗      g 
Sets the upper and lower limit values. 
HH: Secondary upper limit value  HI: Upper limit value 
LO: Lower limit value    LL: Secondary lower limit value   
The unit is what is displayed in the weighing mode. Use a unit of 
three digits in A&D standard format. 
Command e.g.: HI:100.00000    g  
(To set the upper limit value to 100 g:    indicates a space.) 
?HH 
?HI 
?LO 
?LL 
Outputs the upper and lower limit values. 
?HH: Secondary upper limit value   ?HI: Upper limit value 
?LO: Lower limit value  ?LL: Secondary lower limit value  
?HI response e.g.: HI,+100 .00000    g  
 
Commands to query memory data 
MCL 
Deletes all weighing data in memory. 
MD:nnn 
Deletes weighing data with the data number nnn. 
?MA 
Outputs all weighing data in memory. 
?MQnnn 
Outputs weighing data with the data number nnn. 
?MX 
Outputs the number of weighing data in memory. 
UN:mm 
Recalls the unit mass stored in memory with the number of mm (01-20). 
?UN 
Outputs the unit mass number of the selected unit mass. 
?UN response e.g.: UN,01 
CN:mm 
Changes to the upper/lower limit value in memory with the number of mm. 
(mm is the upper/lower limit value code number. Standard=01 to 20) 
?CN 
Outputs the upper/lower limit value code number of the selected limit value. 
Notes:  nnn indicates a three-digit numerical value. 
 
The command is sent to the balance with the terminator selected in the “Terminator 
(Crlf)” parameter of “Serial interface ( 5if )” added.